Under SFDA guidance MDS-G008, a device’s class follows from its intended use, and the SFDA reads that intended use from your advertising as well as from your technical file.

All insights · Published 26 November 2026

Who classifies the device?

The manufacturer does. The Medical Devices Law says the manufacturer shall classify medical devices in accordance with the classification system, and the rules are set out in Annex 5 of the marketing authorization requirements, MDS-REQ 1. Guidance on Medical Devices Classification, MDS-G008, explains how to apply them.

The class shapes the Medical Devices Marketing Authorization (MDMA) application. For low-risk devices, other than IVDs, novel devices and the special Class A cases below, MDS-REQ 1 asks for a shorter set of technical documents.

What are the classes?

ClassLevel of risk
Class ALow
Class A, supplied sterileLow–medium
Class A, with a measuring functionLow–medium
Class A, reusable surgical instrumentsLow–medium
Class BLow–medium
Class CMedium–high
Class DHigh

IVD devices are classified with separate rules into four classes, from Class A (low individual risk and low public health risk) to Class D (high individual risk and high public health risk).

Why do your marketing claims matter?

The class depends on the manufacturer’s intended use, the risk to patients and users, how invasive the device is and how long it is used. MDS-G008 notes that identical devices can fall into different classes if they are used in different parts of the body, which is why the intended use is critical. The SFDA reads it from four sources:

  • The instructions for use (IFU)
  • The label
  • The manufacturer’s advertising materials
  • The technical documentation

So a brochure that claims more than your IFU can point to a different class, and a different file. The guidance puts the onus on the manufacturer: to avoid a higher class, the intended purpose must be defined clearly in the labelling, with positive or negative indications for use. Its example is a catheter that could be used in the central circulatory system, where an explicit exclusion in the intended use keeps it out of the higher class.

Advertising for a medical device also needs its own SFDA approval before it is published, under MDS-REQ 8.

What happens when more than one rule applies?

There are 22 classification rules for general medical devices. If several rules, or several sub-rules within one rule, apply to the same device, the strictest one, giving the higher class, applies. The example in MDS-G008 is a wound dressing that contains collagen: one rule would place it in Class A, B or C, another in Class D, so it is Class D.

A device that is not intended for one specific part of the body is classified by its most critical specified use. Accessories are classified in their own right, separately from the device they are used with.

How is software classified?

Software that drives a device or influences its use falls in the same class as that device. Stand-alone software is classified in its own right, under rule 11: software that provides information used for diagnostic or therapeutic decisions is Class B, Class C if a decision could cause a serious deterioration of health or a surgical intervention, and Class D if it could cause death or an irreversible deterioration of health.

Check the class first

Classification is the first thing we check on every project, for each device in a portfolio. Getting it right early keeps the file, the timeline and the marketing claims in line with each other.

Insights

More insights

SFDA approval for medical device advertising (MDS-REQ 8)

The SFDA rules for medical device advertising under MDS-REQ 8: prior approval, language, websites, social media and content creators, review times, fees and validity.

Read the article

SFDA requirements for AI medical devices (MDS-G010)

What SFDA guidance MDS-G010 asks of AI and machine learning medical devices: clinical evaluation, data, risk management, quality management and change notification, and why the reliance pathway may not apply.

Read the article

SFDA reliance pathway (MDS-G30) for medical devices

How the SFDA reliance pathway works under MDS-G30: the six recognized regulators, who qualifies, what is excluded, how to apply and the 35–60 working day review.

Read the article
Device classification

Not sure of a device’s class?

Send us the IFU and the label. We’ll check the class under SFDA rules before you build the file.

Talk to a specialistinfo@i-regulations.com
+966 53 530 0129