Approved elsewhere? The SFDA’s reliance pathway, explained.

Since September 2026, an MDMA application can rely on an approval from one of six recognized regulators, so the SFDA does not have to assess the device from scratch. Here is who qualifies, what is excluded and how to apply.

All insights · Published 26 November 2026

What is the reliance pathway?

Reliance means that one regulator takes into account, and gives significant weight to, an assessment another regulator has already made. Under MDS-G30, manufacturers and their Authorized Representatives (ARs) can use a marketing authorization from a recognized regulator to support a Medical Devices Marketing Authorization (MDMA) application. The SFDA’s stated aim is to reduce documentation requirements and shorten review timelines.

The SFDA remains responsible for its own decision. The guidance says it keeps full decision-making authority and may ask for more information during the review.

Which regulators does the SFDA rely on?

Country or jurisdictionRegulatory authority
AustraliaTherapeutic Goods Administration (TGA)
CanadaHealth Canada
European UnionNational competent authorities of the European Economic Area (EEA)
JapanPharmaceuticals and Medical Devices Agency (PMDA)
United KingdomMedicines and Healthcare products Regulatory Agency (MHRA)
United StatesU.S. Food and Drug Administration (FDA)

The SFDA may add jurisdictions over time, including other members of the International Medical Device Regulators Forum (IMDRF) Management Committee.

Who qualifies?

An MDMA application is eligible for the reliance pathway when it meets all three criteria:

  1. The device holds a valid marketing authorization or approval from at least one RRA.
  2. It is technically identical to the device the RRA authorized, including its design, manufacturing process and intended use.
  3. It meets the requirements specific to Saudi Arabia: for example electrical power voltage, frequency and conductor colours under Part 401 of the Saudi Building Code; the content and language of the labelling, instructions for use (IFU) and user interface; and storage and transport conditions for temperature and humidity.

For the quality management system, the SFDA accepts Medical Device Single Audit Program (MDSAP) certificates and audit reports from MDSAP auditing organizations as evidence of conformity.

What is outside the pathway?

  • Self-registration routes: products on the market through a manufacturer’s self-registration or self-declaration, without an independent assessment by a regulator or a recognized third-party conformity assessment body.
  • Equivalence to a predicate: products granted market access mainly by showing substantial equivalence to a predicate device.
  • Custom-made medical devices.
  • Products with post-market safety concerns: those subject to significant recalls, suspensions, withdrawals or restrictions imposed by an RRA.
  • Software as a Medical Device (SaMD) and AI-enabled medical devices, which the SFDA may exclude based on their intended use, risk, novelty, complexity or other considerations.

The predicate exclusion is the one to check early. For a device with US market access, how that access was obtained matters, not only the fact that it was obtained.

How do you apply?

Reliance is chosen inside the normal MDMA application on the SFDA e-services portal. The guidance sets out three steps:

  1. Create an account and obtain an establishment number, unless you already have one.
  2. In the Overview section of the MDMA application, answer “Yes” to the question in subsection 2.1, then give the country and title of each RRA approval and upload the evidence.
  3. Complete the Reliance Form in Annex 1 of MDS-G30, printed on the legal manufacturer’s letterhead and signed, and upload it named “Reliance Form”, with Saudi Arabia selected as the country.

The Reliance Form asks for a rationale: the device’s worldwide regulatory status, the evidence that reliance is appropriate, and confirmation that the device is identical to the one the RRA authorized, including its intended purpose, design, specifications, manufacturing information and labelling.

The application must also follow the SFDA classification rules in MDS-G008 and, where several devices share one application, the bundling criteria in MDS-G028. For novel, complex or higher-risk products, such as AI-enabled devices, companion diagnostics and combination products, the SFDA may ask for more information.

How long does the review take?

The standard review for a reliance application is 35–60 working days, for Class A, B, C and D devices alike, provided the application is complete. The guidance says review time may be reduced to 35 working days, subject to eligibility and complete documentation.

Reliance continues after approval. The SFDA also draws on post-market decisions and safety information from RRAs, such as recalls and field safety corrective actions, when it decides whether action is needed in the Saudi market.

Where to start

Gather your RRA approvals, check each device against the exclusions and the Saudi-specific requirements, and confirm its SFDA class. As Authorized Representative, IREG prepares and submits MDMA applications for overseas manufacturers and can tell you which of your devices fit the reliance pathway.

Sources

This article summarises the SFDA documents listed above as they stood on the day it was last reviewed. It is general information, not advice on a specific product. SFDA documents change, so check the current version on the SFDA website, or ask us.

Related services: Overseas manufacturers · Medical devices

Insights

More insights

SFDA post-market reporting deadlines (MDS-REQ 11)

The SFDA post-market surveillance deadlines under MDS-REQ 11: adverse events in 2, 10 or 30 days, final reports in 15, 30 or 60 days, and field safety corrective actions in 2 days.

Read the article

MDMA vs MDEL: the two SFDA medical device licences

The difference between the SFDA’s Medical Devices Marketing Authorization (MDMA) for a device and the establishment licence (MDEL) for the companies that represent, import and distribute it.

Read the article

SFDA medical device classification (MDS-G008)

How the SFDA classifies medical devices under MDS-G008: classes A to D, why the intended use is read from your label, IFU and advertising, and the rule that the highest class applies.

Read the article
Reliance pathway

Send us your approvals.

We’ll map the pathway for each of your devices, whether it is reliance or a full review.

Talk to a specialistinfo@i-regulations.com
+966 53 530 0129