The MDMA starts a clock: SFDA reporting deadlines.

Once a device is on the Saudi market, the SFDA’s post-market surveillance requirements set fixed deadlines for reporting adverse events and field safety corrective actions. Here they are in one place.

All insights · Published 26 November 2026

Who has to report?

Requirements for Post-Market Surveillance of Medical Devices, MDS-REQ 11, applies across the supply chain:

  • Manufacturers, Authorized Representatives and healthcare providers report adverse events and complaints to the NCMDR, and follow up the investigation.
  • Manufacturers, Authorized Representatives, importers and distributors each appoint an authorized person to communicate with the SFDA, keep written procedures for following up incidents, and run a tracking system for the supply and distribution of their devices.
  • Healthcare providers appoint a contact officer with the NCMDR.

Reports go through the NCMDR, Saudi Vigilance or the call centre on 19999.

How fast must an adverse event be reported?

DeadlineWhen the adverse event or complaint
2 daysRepresents a serious public health threat
10 daysRepresents a threat that may cause or contribute, directly or indirectly, to death or serious injury
30 daysHas any other effect

Each deadline counts from the date the event occurred or the date you became aware of it. When the NCMDR contacts you to follow up an investigation, you have 5 days to respond.

When is the final report due?

Investigation reports come in three stages: an initial report within the deadlines above, follow-up reports as the investigation progresses, and a final report setting out the actions taken and the recommendations. The final report is due within:

DeadlineInvestigation
15 daysNo testing or technical evaluation is needed
30 daysThe device is tested inside Saudi Arabia
60 daysThe device is tested outside Saudi Arabia

These deadlines also count from the date of occurrence or awareness. If an investigation takes more than 30 days, submit a follow-up report with a justification, which the SFDA assesses.

What are the deadlines for a field safety corrective action?

Manufacturers and Authorized Representatives report to the NCMDR any FSCA or warning that affects the Kingdom, whether it was issued by the manufacturer or by a regulator abroad. MDS-REQ 11 sets the following deadlines:

DeadlineStep
2 daysReport the FSCA to the NCMDR, with the FSCA letter, counted from the date of the letter
5 daysRespond to an SFDA safety alert or an NCMDR inquiry about an FSCA
5 daysNotify importers, distributors, healthcare providers and users, counted from reporting to the NCMDR
5 daysSubmit the FSCA implementation plan, counted from reporting to the NCMDR
Over 90 daysIf implementation will take this long, the plan sets dates for periodic progress reports

Notification must be documented, for example with signed acknowledgement letters, and the records must show at least three attempts by two different methods. Closure counts only once the NCMDR confirms it.

Other deadlines worth knowing

  • Violating devices: a corrective plan for fraudulent, unregistered or unauthorized devices is due within 5 days of reporting them to the SFDA, and needs SFDA approval.
  • Changes to an MDMA: under the marketing authorization requirements, MDS-REQ 1, the SFDA is informed of a significant change within 10 days and of a non-significant change within 30 days.

Make each deadline someone’s job

The deadlines are short, so decide before launch who receives complaints, who communicates with the NCMDR and who holds the distribution records. As Authorized Representative, IREG reports recalls, field safety corrective actions and adverse events to the NCMDR for the manufacturers it represents.

Insights

More insights

MDMA vs MDEL: the two SFDA medical device licences

The difference between the SFDA’s Medical Devices Marketing Authorization (MDMA) for a device and the establishment licence (MDEL) for the companies that represent, import and distribute it.

Read the article

SFDA medical device classification (MDS-G008)

How the SFDA classifies medical devices under MDS-G008: classes A to D, why the intended use is read from your label, IFU and advertising, and the rule that the highest class applies.

Read the article

SFDA approval for medical device advertising (MDS-REQ 8)

The SFDA rules for medical device advertising under MDS-REQ 8: prior approval, language, websites, social media and content creators, review times, fees and validity.

Read the article
Post-market surveillance

Map who owns each deadline.

Tell us about your devices and your Saudi distributors. We’ll set out who reports what, and when.

Talk to a specialistinfo@i-regulations.com
+966 53 530 0129